Transmitter CC22
Detection of combustible gases
Superior technology
The CC22 works according to the heat tone principle, a proven method for measuring combustible gases. The transmitters are fitted with high-performance electronics for voltage stabilization, processing of readings and temperature compensation. Follow-up costs are minimized thanks to the long service life of the sensors.
Signal processing
The embedded software of the CC22 linearises the measuring signal and compensates for the temperature impacts. As a result, correct readings are transmitted even with weather-related variations in temperature. In addition, the transmitter can be connected to a gas measurement computer for transmitting service, maintenance and error messages.
For transmitting the signals, the CC22 either uses an analogue (0.2 .. 1 mA / 4 .. 20 mA) or a digital (RS485) interface. This permits the universal use of the transmitters in combination with gas measurement computers of GfG and thus with your safety system such that you and your systems will benefit from the best possible protection.
Variants for each use
CC22 in basic configuration
CC22 D with additional indication of the current gas concentration, visual and audible alarm directly on the display and calibration through the keyboard
The benefits at a glance
- Gas concentration displayed on transmitter
- Few cabling required
- Long service life of sensors
- Low maintenance effort
- Status and function display directly on transmitter
- Calibration by pressing a button without having to open the casing
- On-site alarm through coloured display and integrated audible alarm
Technical data
- Measurement gases: combustible gases and vapours, e.g. CH4, C3H8, H2
- Measurement ranges: 0 .. 100 % UEG
- Measurement principle: Heat tone
- Gas supply: Diffusion or gassing with calibration adapter
- Response time t90: less than 30 seconds (depending on gas type)
- Casing: Plastic
- Weight: up to 325 g (depending on variant)
- Dimensions: 98 x 120 x 50 mm
- Protection type: IP54
- Expected average service life of measuring cell: Guarantee > 5 years
- Ambient temperature: -20°C .. +50°C
- Relative humidity: 5 .. 90 % r. h.
- Pressure: 950 .. 1100 hPa
- Output signal: 0.2 .. 1 mA / 4 .. 20 mA / RS485
- Power supply: 18 .. 30 Volts DC